← Search

Imaging · dataset · 2025

Transcranial ultrasound hydrophone scanning and X-ray CT data

Listed in ETH Zürich Research Collection

Experiments were conducted on an ex-vivo human skull (87-year-old male), in compliance with the Swiss Association of Research Ethics Committees and approved by the Kantonale Ethikkommission Zürich (BASEC-Nr 2022-01238).

Description

The skull was preserved in Thiel’s solution, degassed in saline under vacuum for ≥24 hours, and immersed in water during CT scanning to eliminate air. CT scans were acquired using a Revolution CT (GE MEDICAL SYSTEMS) at 120 kVp, reconstructed with bone window filtering at 391×391×625 µm³ resolution and a 512×512 in-plane matrix.

Transcranial acoustic pressure fields were measured at normal and 40° incidence angles. A dual-laser alignment method ensured accurate angle setup: one laser established the surface normal, and a second was adjusted to the desired angle. The transducer was aligned with the second laser beam, ensuring its acoustic axis formed a 40° angle with the XZ plane, lying entirely in the XY plane (Z=0).

Read the rest (3 more)

Additional laser beams along Y and Z axes confirmed the measurement point was at the skull’s outermost surface, compensating for curvature. Measurements used a focused transducer (Imasonic A101, 0.55 MHz, 80 mm diameter, 64 mm radius of curvature) and a 1 mm PVdF needle hydrophone (Precision Acoustics, UK; 0.1–10 MHz calibrated range). The transducer, driven by a signal generator and ENI 2100 L RF amplifier, emitted 500 kHz pulses at 10 kHz repetition rate.

The –6 dB focal spot diameter was 3 mm. The hydrophone, mounted on a 3-axis computer-controlled system, scanned a 35×35 mm² area in XY and XZ planes. Scanning was continuous along X, with 0.3 mm steps in Y and Z. Spatial averaging was performed over 0.3 mm along X. Scan limits were verified to avoid hydrophone-skull collisions.

Reference scans were performed post-measurement with the skull removed, maintaining identical transducer and hydrophone positions. Measurements were taken at parietal and frontal bone regions, with bone thicknesses of ~5.2 mm and ~6.1 mm, respectively, at normal and 40° incidence. All measurements were conducted in a tank filled with degassed, de-ionized water.

Links

Where it is published

Catalogue records · 1

Topics

Inferred from text
Computed tomography 50%
Provenance · 1 source records, 19 field assertions
SourceKeyLast seenRaw
ETH Zürich Research Collectionoai:www.research-collection.ethz.ch:20.500.11850/78284312 d agoJSON v1
FieldAssertionExtractorEvidence
access_levelsource · research collection ethz chconnector:research_collection_ethz_ch@1.0.0
concepts[field].local:field:chemistrymapping · research collection ethz chconnector:research_collection_ethz_ch@1.0.0
concepts[field].local:field:computer-science-aimapping · research collection ethz chconnector:research_collection_ethz_ch@1.0.0
concepts[field].local:field:earth-environmentalmapping · research collection ethz chconnector:research_collection_ethz_ch@1.0.0
concepts[field].local:field:economics-financemapping · research collection ethz chconnector:research_collection_ethz_ch@1.0.0
concepts[field].local:field:engineeringmapping · research collection ethz chconnector:research_collection_ethz_ch@1.0.0
concepts[field].local:field:humanitiesmapping · research collection ethz chconnector:research_collection_ethz_ch@1.0.0
concepts[field].local:field:life-sciencesmapping · research collection ethz chconnector:research_collection_ethz_ch@1.0.0
concepts[field].local:field:materials-sciencemapping · research collection ethz chconnector:research_collection_ethz_ch@1.0.0
concepts[field].local:field:mathematics-statisticsmapping · research collection ethz chconnector:research_collection_ethz_ch@1.0.0
concepts[field].local:field:medicine-healthmapping · research collection ethz chconnector:research_collection_ethz_ch@1.0.0
concepts[field].local:field:physicsmapping · research collection ethz chconnector:research_collection_ethz_ch@1.0.0
concepts[field].local:field:psychology-behavioralmapping · research collection ethz chconnector:research_collection_ethz_ch@1.0.0
concepts[field].local:field:social-sciencemapping · research collection ethz chconnector:research_collection_ethz_ch@1.0.0
concepts[modality].local:modality:ctenrichment · research collection ethz chkeyword-concept-rules@1.0.0title+description (50%)
descriptionsource · research collection ethz chconnector:research_collection_ethz_ch@1.0.0/metadata/dc/description
license_textsource · research collection ethz chconnector:research_collection_ethz_ch@1.0.0
publication_datesource · research collection ethz chconnector:research_collection_ethz_ch@1.0.0
titlesource · research collection ethz chconnector:research_collection_ethz_ch@1.0.0/metadata/dc/title